Automatic Single Syringe Pump with High Quality

Model NO.: SP-50
Origin: Beijing, China
CHARACTERISTICS

Compatible with Syringes of any Brands
Automatic calibration for Syringes of any Brand.
Three working modes: Rate control mode, Time control mode and Dose-Weight control mode.
Automatic Detection of Syringe types and settings.
Broad range in delivery rate settings
8 Hours internal battery life in high capacity.
Three levels of alarm volume: High, medium and low.
Three levels of occlusion pressure: High, medium and low.
KVO function: After infusion is complete, the pump automatically switches to the smallest flow to prevent creating the thrombus.
SPECIFICATIONS

Compatible Syringe: Any standard syringe with 10ml, 20ml(30ml) and 50ml(60ml)
Accuracy: ± 2% (Any manufacturers of syringes, right after calibration )
Delivery Rate Range:
50ml(60ml): 0.1ml/h~1800.0ml/h (0.1ml/h step)
20ml(30ml): 0.1ml/h~800.0ml/h (0.1ml/h step)
10ml: 0.1ml/h~400.0ml/h (0.1ml/h step)
Time Mode Range:
Volume: 0.1~999.9ml (0.1ml step)
TIME: 1~1999min (1min step)
Body Weight Mode Range:
Volume: 0.1~999.9ml(0.1ml step)
DRUG MASS: 0.1~999.9mg(0.1mg step)
DOSE: 0.01~99.99ug/kg/min(0.01ug/kg/min step)
WEIGHT: 0.1~300kg(0.1kg the least step)
Volume Limit Range: 0~1999.9ml (0.1 ml step)
Total Volume Range: 0.1~1999.9ml (0.1 ml step)
Rapid injection: 50ml(60ml): 1800.0ml/h, 20ml(30ml): 800.0ml/h, 10ml: 400.0ml/h
KVO Rate: 1ml/h
Alarms:
Near empty, Syringe empty, Volume complete(a volume limit has been set ), Occlusion, Syringe loose, Operation error, Low battery and other device failures.
Power Supply:
AC power supply: AC 100 to 240V± 10%, 50/60Hz;
Internal LI Battery: 11.1V, 2000mAh. Approximately 8 hours of continuous operation on a new and fully charged battery running at 5ml/h.
Power Consumption: ≤ 25VA
Safety level: Class I, Type BF
Dimensions: 280mm(L)× 160mm(W)× 130mm(H)
Net Weight: Approximately. 3.0kg
Operating Conditions: Ambient temperature: 5 to 40 Relative humidity: 20 to 90%
Storage Condition: Ambient temperature: -10 to 55 Relative humidity: 10 to 90%
Accessories: AC power line, Operational manual
Flow Rate 1ml/h~1300ml/h
More than 1000ml/h, 1ml/h step
Less than 1000ml/h, 0.1ml/h step
Standard IV SET ±5%
Preset infusion volume 1ml~9999ml
More than 1000ml, 1ml step
Less than 1000ml, 0.1ml step
Occlusion sensitivity Optional choose from higher to lower
KVO rate 1 ~5ml/h (adjustable)
Power supply 110~220 + 10%, 50~60Hz + 1Hz
Power consumption Less than 20VA
Battery Life More than 3 hours continuously at 30ml/h with fully charged battery
Work Condition Circumstance Temperature: 10 ~40 degree infusion precision will decrease when temperature ≤18 degree Relative Humidity: 30~93% Atmospheric Pressure: 86~106KPa
Pump dimension (mm) 233 x146x269
Pump weight About 3kg
Applicable scope Fit for mainline, small volume infusion, quick infusion